Laboratory Balances
|
by Frank Weithöner
|
Laboratory balances or micro-balances are used in hospital laboratories, pharmaceutical industry, research institutions and
schools. They are designed to determine precisely very small weights. Resolutions of a thousandth of a gram are not unusual.
Laboratory balances are sophisticated and sensitive electromechanical devices which need special care and knowledge in the
usage, maintenance and repair.
|

Laboratory balances for different purposes and requirements.
|
In laboratories we usually find two types of balances: The precision balance and the analytical balance. The difference is the
accuracy. Analytical balances are more precise and thus more expensive.
But progress in electronics in the last years have simplified the usage and made balances cheaper. Precision balances are
available from 400 €, analytical balances from 1,500 €.
Balance or scale?
Both, the term balance and weighing scale are used when people talk about the weighing of masses. Both expressions can be used
but in practice the following definition prevails: Scales are used for weighing larger masses like a bathroom scale or baby scale
and balances are used for precise weighing of substances, as such are used in laboratories.
Weight or mass?
A bit more tricky is the difference between weight and mass. Both words are used interchangeably. Strictly speaking we must
consider the following facts:

The unit of the mass is the kilogram [kg]. It
is equal to the mass of the international
prototype kilogram and therefore fixed.

The weight is a force which is expressed in the unit Newton [N].
It is the product of mass
and the earth's gravity. Therefore,
the value is not fixed but depending on the location.
This is important when we think about the different weighing principles. A spring scale mechanic of a bathroom scale for example
measures the distance a spring deflects under its load. It is influenced by the altitude (force of gravity) and so it measures the
weight.
A laboratory balance measures a
mass by comparing a unknown mass with a reference mass.
Unfortunately we always say, that we measure a weight and use calibration weights which is strictly spoken not correct: A balance
measures a mass.
Technical expressions
When we talk about the balance's specification and measurement features we will stumble over technical expressions which may not
necessarily be clear.
Here are some terms we should know:
Accuracy How close the balance display is to the actual weight of a weighing
sample.
Adjustment Calibration
Calibration The deviation between the measurement value showing in the display and the true value of the measurement. An
adjustment with internal or external calibration weights should then be made.
Capacity The largest weight the balance is capable of measuring.
Corner load The deviation of the measurement value through off-center loading.
Digit The smallest increment of weight that the digital display resolves.
Drift The drift is the slow change of the readout over a long time and with a constant load on the balance.
Linearity Deviation from the theoretical linear curve of two variables: The zero point and the calibrated point close to
the maximum capacity.
Ppm Parts per million = 10
-6 (e. g. 1 mg of 1 kg)
Readability The precision of a balance. Readability is the smallest difference in mass that can be displayed by the balance.
Repeatability The ability of the balance to produce the same result with one and the same load under the same measurement conditions.
Reproducibility Repeatability
Resolution There is no standardized definition. In general it is used to indicate the readability compare to the maximum capacity of
the balance. The resolution is the quotient of maximum capacity and readability.
Example: An analytical balance with a weighing capacity of 200 g and a readability of 0.1 mg has a resolution of 2,000,000 digits.
Response time Stabilization time
Stabilization time The time it takes from placing the sample on the balance and displaying the final result of the measurement.
Taring This function sets the display to zero when a load (e.g. an empty container) is already on the weighing pan. The net weight
will be displayed then, after the container has been filled.
Weighing range Capacity
Types of balances
In laboratories we usually find two different types of balances: The precision balance and analytical balance. The difference is the accuracy. Analytical balances are more precise and more expensive.
Semi-micro balances and micro balances are even more precise but not very common in hospital laboratories.
Here the standard readabilities for various balance types:
Type
|
Readability
|
Precision balance
|
1 mg
|
0.001 g
|
or
|
3 digit
|
Analytical balance
|
0.1 mg
|
0.0001 g
|
or
|
4 digit
|
Semi-micro balance
|
0.01 mg
|
0.00001 g
|
or
|
5 digit
|
Micro balance
|
1 µg
|
0.000001 g
|
or
|
6 digit
|
Precision balances
Precision balances are used to measure mass with a precision of up to 1 mg. They are widely used because their accuracy is high enough for the common
measurements.
|

Precision balances are highly reliable, robust and compact.
|
Precision balances are nowadays all microprocessor controlled. That makes the usage easy and brings more functionality. Beside normal
and tare weighing, additional features are: piece counting, recipe weighing, percentage weighing, change of weighing units and
the connection with other equipment or with a computer via data interface.
In principle all balances run on low-voltage or battery which is important for laboratories with unreliable power supply.
Most of the precision balances have no internal calibration weights. For calibration certified calibration weights of class E2 or better E1
are needed.
Precision balances come in different capacities. Balances of 50 g capacity are available as well as for 10 kg capacity. The
readability of a typical precision balance of 100 g is 1 mg. A 1-kg-balance has a typical readability of 10 mg and a balance for
up to 10 kg 50 mg.
The price for a common precision balance is between 100 and 500 €.
Analytical balances
Analytical balances are found in laboratories for the manufacture of pharmaceutical products, analysis, quality assurance or
final checking. They are more precise, sensitive and more expensive. With a readability of 0.1 mg they are ten times more precise
than precision balances.
|

Analytical balances are covered with a glass enclosure in order to limit measurement inaccuracy caused by dust and air flow.
|
Due to the high resolution analytical balances are also more sensitive to outside influences like dust and air flow. To avoid these
factors, the weighing pan of analytical balances is protected by a transparent enclosure with sliding doors.
The electronic control units of precision and analytical balances are in principle identical so that the application possibilities
are the same. Also analytical balances offer piece counting, recipe weighing, percentage weighing, change of weighing units and a data
interface.
The only difference is an additional function: The automated calibration function. That means, that no external calibration weights are
needed to calibrate the balance before the measurement. A simple press of the calibration button is enough and a built-in motor-driven
adjusting weight will do the calibration.
This function is an enormous advantage over common balances. The calibration procedure is easier and faster and external weights are
not needed and so can not get damaged.
The price for analytical balances start from 800 €.
Using laboratory balances
Thanks to modern electronics, the handling of laboratory balances has become easier and faster. Nevertheless, the user manual should be
read by both the operator and the technician. The usage varies from manufacturer to manufacturer and navigating through the menu and
additional functions may not be clear immediately. Only who has read the manual knows about all functions and the right working procedures
and can avoid mistakes.
Preparations
Laboratory balances are sensitive instruments and a reliable measurement can only be expected when the balance is prepared correctly.

Is the balance exactly levelled? Check the internal spirit level.

Was the balance connected to the mains for at least 4 hours?

Is the balance and the workplace absolutely clean?

Do you know the maximum load? Never overload a balance.
Using procedures
This is just a short overview. By all means, read the user manual of your balance before.

Do a calibration if needed.

Press the tare key to zero the display.

Weigh powders only on paper or small containers.

Never touch samples or samples containers with your bare fingers. Wear gloves or use
anti-magnetic forceps.

Place the sample in the centre of the weighing pan.

Close the enclosure before starting the measurement.

Wait until the stability indicator is displayed.

Have you spilled chemicals on the balance? Clean up immediately.
Tare-Function
When a load is already on the balance, pressing the tare-button sets the display to zero. It is used when a samples comes in a container.
First the empty container is measured, the tare-button is pressed and then the filled container is placed back. The result is the net weight.
Warm-up time
A laboratory balance should always be connected to the power. Don't disconnect the balance from the wall socket or power supply. Switch off the
balance only with the key of the keypad. Even when it is switched off and the balance is in standby mode, the measurement unit is still powered
and has the necessary operating temperature. The balance then can be used immediately after switching on.
If the balance was disconnected from the mains the balance has to warm up for at least 4 hours before the first weighing. Only then is the
measurement accurate and reproducible.
Cleaning
After every usage and at the end of the day the balance has to be cleaned carefully.

Keep the weighing pan and weighing chamber clean at all times.

Use a fine paintbrush to remove sample residues. If possible remove the weighing pan for
cleaning.

Use an absorbent cloth to remove spilled liquids.

If there is stain use a damp cloth and a mild soap solution for cleaning. Do not make the
cloth too wet. Make sure that no moisture enters the balance. Wipe the balance with a
soft and dry cloth afterwards.

The glass of the weighing chamber of an analytical balance can be cleaned with a
common window cleaner.
Tip! Buy a dust cover for your balance. It is cheap and helps to protect your
expensive balance against dust and dirt.
|
User manuals
User manuals are always delivered with the balance. It is wise to copy the manual and leave one issue close to the equipment and keep one in
a file with other technical manuals in the office or storage room.
If the manual is missing, you are probably able to download it from the manufacturers website. If this is not possible, try it
here:
Calibration
The more precise the measurement result has to be, the more often a calibration has to be done. While a weekly calibration for a normal precision
balance is enough, for analytical balances a calibration before every measurement is common.
In order to keep the the calibration simple for the operator, analytical balances usually contain an automated calibration function. Internal
calibration weights are then automatically placed on the sensor. The user only has to press a button and the balance calibrates itself. For all
other balances extra calibration weights have to be taken and the user has to place them on the weighing plate manually.
Calibration does not mean, that a technician does adjustments to the inside of the equipment. Calibration means checking the correctness of the
display by comparing the read-out with a calibrated reference weight by the user.
In order to calibrate a balance, we have to control and adjust two calibration points: Zero and full load. For zero we only reset the display when
the balance is unloaded. For determine the full load point we need a precise calibration weight which is close to the capacity of the balance.
Once these two extremities are set, the balance will now be able to calculate all quantities between these two points.
While the laboratory assistance does the calibration when ever he needs it, the biomedical technician is responsible for calibration during the
installation and after every repair. The calibration always has to be done at the workplace in the laboratory and after every relocation. It is
also advisable to make a frequent function check and calibration after some time. The sensor and the electronics 'ages' and the measurement
drifts over time.
Preparations
Before calibrating a balance, we have to make sure, that all influences from the outside are excluded. Interfering factors are:

Unleveled balance or uneven surface of lab bench

Air flow

Unstable lab temperature

Steel lab bench or equipment which produce magnetic fields close to the balance

Interrupted power supply
Do the calibration according the user manual. Every balance is different and a general instruction does not exist. Only there the correct handling
is explained.
If you want to calibrate a balance without an automated calibration function make sure that you have an appropriate calibration weight. The accuracy
of the weight must be much higher than of the balance and the mass must be close to the maximum weighing capacity of the balance.
Calibration with internal calibration weights
Modern analytical balances already have integrated calibration weights. Only a single key has to be pressed and the balance does the whole calibration
process automatically.
First the balance is set to zero without any load on the sensor and then a mechanic places automatically an internal calibration weight on the sensor.
Calibration weights
The balance manufacturers provide sets of calibration weights in various sizes and different calibration weight classes. Which weight class is needed
depends on the accuracy of the balance.
Calibration weights are extremely precise and have to be handled with absolute care. The smallest scratches makes them useless and even dust and
fingerprints distort the calibration. Touch the weights only with clean gloves or with forceps and put them immediately back into the box when the
calibration is finished.
Before doing the calibration, be sure that the balance is in warm operating condition. The balance has to be connected to the power supply for at
least four hours, some manufacturers even suggest ten hours. Only then it is guaranteed that the sensor and the electronics has operation temperature
and does not drift any more.
That does not mean, that the balance has to be switched on four several hours, but it has to be connected to mains and be in stand-by mode. This is
to be noted specially in countries with unreliable power supply where the equipment is always disconnected from mains when not in use.
Weight classes
Calibration weights are available in different weight classes. The weight class describes the accuracy of the weight. For calibrating a precise
analytical balance, a more accurate weight is needed than for calibrating a less accurate balance.
It should be also clear, that the tolerance (quality) of the calibration weights has to be much better than the balance which has to be
calibrated.
These weights are our standard and with the calibration we set our balance to this standard. A calibration is only as good as the calibration
weight.
Important! The tolerance of the weights must be much better than the
resolution of the balance. Otherwise the calibration is useless!
|
Here are the specifications of the different classes of calibration weights:
Class E1
|
Maximum error at 100 g
|
± 0.1 mg
|
199.9999 g - 200.0001 g
|
Maximum error at 1 kg
|
± 0.5 mg
|
999.9995 g - 10.0005 g
|
Class E2
|
Maximum error at 100 g
|
± 0.3 mg
|
199.9997 g - 200.0003 g
|
Maximum error at 1 kg
|
± 1.5 mg
|
999.9985 g - 10.0015 g
|
Class F1
|
Maximum error at 100 g
|
± 1 mg
|
199.999 g - 200.001 g
|
Maximum error at 1 kg
|
± 5 mg
|
999.995 g - 10.005 g
|
Class F2
|
Maximum error at 100 g
|
± 3 mg
|
199.997 g - 200.003 g
|
Maximum error at 1 kg
|
± 15 mg
|
999.985 g - 10.015 g
|
Class M1
|
Maximum error at 100 g
|
± 10 mg
|
199.99 g - 200.01 g
|
Maximum error at 1 kg
|
± 50 mg
|
999.95 g - 10.05 g
|
Calibration without internal calibration weights
Laboratory balances without internal calibration weights need external weights. The calibration has to be done manually by the
user. That takes longer and the user has to work very carefully. But this does
not mean, that a manual calibration is less
precise.
Hint! Some balances without internal weights have a 'Calibration'
key. By pressing this key the balance is just set to zero. The calibration with a weight still has to be done manually.
|

Check out the capacity of the balance. Take the information from the balance or the user
manual.

Choose the appropriate calibration weight.
Example: Is the capacity 200 g, the weight also should have 200 g.

Check out the accuracy of the balance you want to test. Take the information from
balance or the user manual. The numbers of digits of the display are meaningless.

Take appropriate calibration weights which are much more precise than the balance.
Example: Has the balance a accuracy of ± 1.0 mg, Class-E1-weights are fine (± 0.1 mg).

Do the calibration according to the user manual.

Calculate the permitted tolerances of the balance according the accuracy taken from the
manual and check if the read out is within the calculated limits.
Example: A 200 g-weight is displayed with 200.00178
The accuracy of the balance is only 1.0 mg. We have to round the result to 200.002
The calculated result must be between 199.999 g - 200.001 g (± 1.0 mg)
The tolerance is too big. The balance has to be calibrated to 200.000 g.
Calibration weights and analytical balances
In the above mentioned example, the laboratory balance with an accuracy of ± 1.0 mg was not a very precise one. Standard analytical
balances have at least a accuracy of ± 0.1 mg and better ones even ± 0.01 mg. If we now want to calibrate an analytical balance we
have a problem: We need calibration weights with a much smaller tolerance than the accuracy of the analytical balance. But how is
that possible, when the most precise calibration weight we can get, only has a tolerance of ± 0.1 mg (Class E1). The answer is easy:
It is not possible. The calibration of an analytical balance has only to be made with internal calibration weights. All existing
external calibration weights are not precise enough.
Important! Never use external calibration weights to calibrate analytical
balances! An analytical balance is more precise than the weights.
|
Tip! Why not using an analytical balance to calibrate the calibration
weights? Even a set of less accurate and cheaper low class weights will give a highly reference for future measurements on other
balances after a calibration with an analytical balance.
Buy a (set of) calibration weight(s) together with a new balance, determine the calibration weights with the new balance and you have
a nice reference for future calibrations.
|
Installation
The location of the balance has to be chosen very carefully. Only a proper set-up of the balance insures perfect readouts over a long
period of time.
Before setting up the balance we have to make sure that the workplace for the balance is suitable. The following criteria have to be
noted:

The room must be dust free and absolute clean. Dirt and dust will influence the weighing
result and also can causes problems with the mechanic of the balance.

The lab table or bench where the balance is placed, has to be absolutely stable and must
not wobble. No vibrations should influence the measurement. The weighing bench should
be fixed either on the wall or on the floor (or better both). It is a good idea to place the
balance directly over the legs of the lab bench in order to maximize the stability.

The surface of the lab bench has to be flat and the balance has to be aligned absolute
horizontally. Use the internal or an external spirit level.

The surface of the lab bench must not made out of steel. Steel can get magnetize and will
influence the weighing sensor (which is technically a magnetic sensor).

The same goes for any items which produce magnetic fields, like transformer or
loudspeakers. Keep them away from the balance.

Use the weighing bench only for weighing. A centrifuge has no place around a balance. It
creates vibrations and a magnetic field through the motor.

The room temperature has to be kept as constant as possible. Even small temperature
changes can cause drifts in the measurement results.

Remember that also direct sunlight and strong light bulbs will heat up the balance.

Check also if the humidity in the lab. It should not change much and has to be within the
range of the manufacturer's limits. Specially in tropical countries this may lead to a
problem.

Avoid a place of much air flow. Do not place the balance next to a door. Ideally the room
should have sliding doors in order to avoid air flow while opening and closing.

The power supply must be stable and free of power cuts. If this can not guaranteed an
additional UPS (
Uninterruptible
Power
Supply) has
to be used.
|

A bad example:
The table is not very strong, neither fixed to wall nor to the floor, the table-top is much too thin, made out of metal and the balance is
placed close to a motor-driven test equipment which creates vibrations and magnetic interference fields.
|
When the best location is found, we can place the balance. Therefore we have to make sure that the balance itself is levelled perfectly
horizontal. For this purpose all balances have adjustable equipment feet and usually an integrated spirit level. This so called 'bull's
eye level' is a small circular glass device. An air bubble floating in a liquid (spirit) has to be positioned in the centre of the level
by turning the adjustable equipment feet of the balance.
|

The internal spirit level helps to level the balance on the workbench. The screw feet of the balance must be adjusted so that the air
bubble is precisely in the centre of the sight glass of the bubble level.
It is the task of the user to control the correct levelling from time to time but always after the balance has been removed.
|
|

After levelling a new calibration has to be made.
|
|

Load cell balances often have stops which prevent the balance from overloading. Check if they are adjusted well, that there
is space enough for a proper operation and not too much space so that the mechanic can get damaged.
Put the maximum permitted weight on the balance and check the distance.
|
Tip! Consider when buying an expensive balance also to buy a solid
bench for the instrument.
|
Preventive maintenance
The maintenance of balances covers cleaning and some test procedures. Internal adjustments should not be done. If the test
results differ much from the balance's specifications, it is advisable to ship the balance to the manufacturer.
The tests should be done on-site in the laboratory. Moving the balance to a workshop is anyway not a good idea. Any moving
can cause additional problems.
Needed tools
Beside the calibration weights for balances without internal calibration weights, no special tools are needed. A proper set
of screwdrivers, the technical manual for the balance and a clean and bright workplace is all we need.
Step 1: Inspection
Before starting any maintenance or repair, start with a short visually inspection:

Is there any contamination or anything else unusual?

Does the doors close/open smoothly?

Do the switches work properly?

Is the zero-point display stable (without a load)?
Make a note of all objections. We come later to these points. First we have to clean the balance.
Step 2: Cleaning

Use a fine paintbrush to remove sample residues. Often the weighing pan can be removed
for easier cleaning.

If there is stain, use a damp cloth and a mild soap solution. Do not make the cloth too
wet. Make sure that no moisture enters the balance.

Wipe the balance with a soft and dry cloth.

The glass of the weighing chamber of an analytical balance can be cleaned with a
common window cleaner.
Step 3: Maintenance
If the sliding doors move sluggishly, take out the doors and clean the rails and the edges of the glass doors. Do not lubricate. Oil or
grease absorb more dirt over time and the function gets worse.
This also applies for the interior of the balance: Do not lubricate anything. Just clean all components from dust and dirt. Any additional
substance interfere with the balance's mechanic.
Important! Never lubricate a balance unless the manufacturer has expressly
indicated it.
|
Step 4: Standard tests
Beside the calibration procedure some tests can be done, in order to verify the accuracy of the balance. The result of these tests
will show the deviation from a perfect balance with no deviation at all. Since such a balance in practice do not exists, there can
be deviations expected. The problem now is to decide, if these differences are within the tolerance of the manufacturer. Therefore
we have to know the specifications of the balance, compare the results and make a decision: Yes, the test results are within the
tolerance, the balance is OK; or no, out-off the tolerance, the balance has to be repaired.
Since the technician is not the operator of the balance and does not necessarily know for what kind of measurements the balance is
used, it is recommended to do the tests, record the test results and deviations and let the operator decide, if the results can be
accepted and the balance is precise enough for future measurements.
The following performance tests can be done:

Reproducibility (Repeatability)

Cornerload

Linearity
The procedures to perform these test are explained below. Also a test chart for each test can be downloaded. The test charts can be
printed out and the test results can be filled in. Later they can handed over to the responsible person.
The tests can be performed semi-annual as part of the preventive maintenance of all the other laboratory equipment.
Are there are significant changes, check again if there are any influences from outside which could create the problem (static,
air draft, warm-up, vibration, etc.) and repeat the test. If the balance still fails the check, the balance has to be send to
the manufacturer for a recalibration.
Reproducibility (Repeatability)
Reproducibility means that a repeated weighing of the same object must deliver the same weight reading. A calibration weight is
measured for ten times and the readings are put into the test chart. After each weighing cycle the balance has to be returned
to zero.
The following Reproducibility Test Chart can be printed out and used as a test report:
|
Reproducibility Test Chart
40 KB
Download
(My standard deviation calculator you find
here.)
|
Cornerload
A cornerload test is performed in order to find out if the balance delivers the same weight reading for a given object regardless
of its position on the weighing pan.
Position a calibration weight at various locations on the weighing pan. The reading should be the same, within a few digits, at
all positions.
Note! A balance which already delivered large deviations during the
reproducibility test will also deliver wrong results also here.
|
The following Cornerload Test Chart can be printed out and used as a test report:
Linearity
Linearity testing verifies the accuracy of the balance at intermediate values of weight. Since a laboratory balance will often be
used to weigh items much smaller than the maximum capacity of the balance, this is an important test.
The following Linearity Test Chart can be printed out and used as a test report:
Regular maintenance
The balance is a sensitive equipment which must operate absolutely correct in order to ensure the quality of the measurement.
Therefore the adjusting and calibration is very important for any precision balance. Only with a frequent calibration the
accuracy of the balance can be guaranteed.
After every repair and maintenance a calibration has to be done.
A full calibration should be done frequently every few year. Therefore appropriate and traceable reference standards have to be
used.
The performance has to be calculated for each measurement range. The end user finally can decide if the balance still meets his
requirements.
Design
Laboratory balances are mechanical balance coupled to a sensor and a processing electronics. The sensor converts the mechanical
force into an electronic signal. This analogue signal is converted into a digital one, processed and displayed.
The mechanic and the sensor are usually designed as a compact mechanical unit which is adjusted to the following electronics.
Mode of operation
For electronic balances two different weight measuring technologies are used:
Strain gauge and
electromagnetic force restoration.
While the strain gauge principle is used for all scales with higher limits and less accuracy like patient and baby scales, high
quality analytical balances are all based on the principle of electromagnetic force restoration.
Strain gauge scales
Also commonly called load cell.
The working principle of the strain gauge is used for patient scales or baby scales where higher weights are measured and the measurement
result within milligrams is not needed.
The unknown weight is measured by using a strain gauge. This is a aluminium beam, which is milled out in the centre. To one end the weighing
pan is attached, the other end is mounted to the base of the balance. A load on the pan creates as force and the aluminium beam gets deformed.
Since at the thinnest part of the beam strain gauges are embedded, these length-sensitive resistors also get deformed and so change their
resistance.
The following amplifier creates a voltage out of the current through the resistances and delivers a measurement voltage which corresponds
to the weight.
|

A standard resistor bridge (also called Wheatstone bridge) with three fixed resistors and a sensor resistor. When the bridge is in
balance the output voltage is zero.
|
|

In practice often all four resistors are length-sensitive. The output voltage and the dynamic is higher and the sensitivity
to disturbances smaller.
|
|

Here the sensor of a platform balance in reality. On the left side the aluminium beam is mounted to the frame. On the right
side of the beam the weighing platform is usually mounted (on the top where the four screw holes are). Note how massiv the whole construction is and how thin the centre part of the
block is. Here the block will bend under load and here the length-sensitive resistors are mounted. The sensors themselves are
embedded in the aluminium block and sealed with plastic.
|
Note that there is nothing to adjust. The calibration is done electronically by the control unit.
Is the working principle based on load cells, then there is not much you can do wrong. The whole unit is build out of one solid
block. On the other hand, there is also nothing to adjust and to repair.
Electromagnetic force restoration balances
Balances based on the electromagnetic force restoration principle work completely different. Here not the force of the load weight
is measured directly but a counteracting force is created which works against the weight. This force is measured when the
weighing pan is in balance. The method allows more precise measurements than the load cell principle.
The weighing pan with the unknown weight is attached to a force coil. The coil is floating in a magnetic field which is created
by an amplifier. The amplifier delivers always the right current to keep the lever in balance, regardless of the weight on the pan.
The information when the lever is balanced is given by a light barrier.
Due to the fact that the needed current for this balancing is proportional to the weight on the pan, the current is also used
to measure the weight. Therefore the current flows through a precision resistor which creates a measurement voltage. An A/D
unit transforms the voltage into a digital signal. The following micro processor with an appropriate software delivers the
measurement result which is finally shown in the display.
Additional sensors like a displacement detector and a temperature sensor are also controlling the amplifier and ensure the
accuracy of the balance.

Here the force coil mechanic of an analytic balance. Usually the mechanic is protected by a heavy steel cover. The black pin in
the centre carries the weighing plate (removed). The weighing mechanic is hidden under the metal disk. The green, small electronic
board on the right contains the LED light barrier. A lever, connected to the weighing mechanic, moves in this sensor.
The left part of the mechanic shows the automated calibration unit. The motor on the left turns the big black gear wheel below
when the calibration function is activated. Then the calibration weight is attached to the weighing mechanic. The calibration
weight consists of the two brass wheels in the front.
Repair
Balances are often damaged by mishandling or contamination from water or chemicals. These problems require cleaning and testing
as described in the Maintenance section. Rarely component have to be replaced.
Laboratory balances consists in principle of two compact and integrated units: The computer controlled electronic unit and the
mechanic with the sensor, often designed as a solid block. The two units are adjusted to each other and thus leave no space for
repairs or adjustments. Only touch or modify them when you are sure what you do. Otherwise take hands off the sensors. You
probably create more problems than you solve.
Is the balance not working at all, you can have a closer look at the power supply.
In case of a more complex problem or inaccuracy the balance should be send to the manufacturer for repair.
|

Be careful with dismantling balances. In this case only the two marked screws are holding the metal cover in place and can be removed.
The four other screws are for adjustments and must not turned!
|
Note! Before dismantling a laboratory balance consult the service
manual.
|
Common problems
Malfunctions caused by sample residues and spilled liquids are typical for laboratory equipment.
Also problems with the power supply can occur.
A typical problem with load cell scales is, that they are sensitive to over-loading. An excessive load can permanently bend the
load cell. Therefore, great care needs to be taken when using a scale with a load cell. The user should have a rough idea of
how much an object weighs before placing it on the weighing pan. Bent load cells are not repairable and have to be exchanged.
The repair is easy to do but sometimes not economic. The spare part is often not much cheaper than a complete new scale.
|

Most of the laboratory balances contain small batteries. The batteries do not run the balance but backup the stored
microprocessor settings. Their lifetime is limited and should be replaced every two years.
|
Needed tools
No special tools are needed. A proper set of screwdrivers, a multimeter maybe, the technical manual for the balance and
a clean and bright workplace is all we need.
Service manuals
While user manuals are always delivered with equipment, service manuals are never provided by the manufacturer. Also on
the websites of the companies you will not find any service manual. The policy is clear: The companies want to do all
repairs by themselves. Only a certified technician who took part in a special training for a
certain equipment gets a service manual and technical support.
If the service manual is needed, you can try it
here:
Power supply
As mentioned above, the possibilities of a repair of a laboratory balance are limited. An exception is, as always, the
power supply. Here is something to do for the technician. In case of malfunctions, we can check the fuse(s), the output
voltage(s) and the electronics in between. It is quite a lot, because here often problems appear.
Laboratory balances almost always have an external power supply (switch mode) and an internal voltage stabilization.
Here the view inside the balance:
1 The input socket with the plug from the external power supply. Plugs and sockets often
have bad contact.
2 The glass fuse.
In switch mode power supplies, also search for small, black plastic fuses. Remember:
Blown fuses often have a cause.
3 A capacitor. Works together with the following coil as a filter for the DC input voltage.
4 Looks like a transformer, but is just a coil for filtering out any AC components of the DC
voltage.
5 Three stabilizers for three different DC output voltages. In this case +12 V, -12 V and
+5 V. They are easy to check. 7805 means Stabilizer for 5 V, 7812 for 12V. Left is input,
centre ground, right 5 V or 12 V output. Negative stabilizers (79xx) are different: Left is
ground, centre is in and right is voltage output. More about stabilizers
here.
6 Capacitors for stabilizing the output voltages.
7 Flat ribbon cable, which leads to the main board. Be careful with these cables. Once
they are broken, they can not be repaired. In such a case, exchange the cable against
normal wire cables.
Manufacturers
Important manufacturer of laboratory balances are:
A&D, Japan
Kern & Sohn, Germany
Mettler Toledo, Switzerland/USA
Ohaus, USA
Radwag, Poland
Sartorius, Germany

Seca
Should be avoided. Prohibited the download of service manuals.
Links and sources
Wikipedia: Weighing scale
Wikipedia: Spring scale
GWP - Proper Weighing with Laboratory Balances
IES Corporation - Calibration Guide of Balances
IES Corporation - Testing Your Laboratory Balance
Mettler Toledo - Weights
Sartorius - Correct Use and Handling of Analytical and Microbalances
Urs Berli - Weighing Principles